Using a moving mirror goniophotometer to test decorative lights.
Evaluating decorative lights using a moving mirror goniophotometer is a little more complicated than testing flood lights, and it normally requires the processes that are outlined in the following paragraphs: In accordance with the instructions provided by the manufacturer, calibrate the goniophotometer once it has been properly installed. This may include positioning the measuring chamber, adjusting the measurement angles, and confirming the accuracy of the photometer or lux meter. Place the decorative light to be examined within the measuring chamber of the goniophotometer. Ensure that the light is placed firmly and that it is aligned with the rotation axis of the goniometer. Adjust the measurement angles on the goniophotometer so that they correspond to the angles given in the standard that you are using. For instance, if you are using ANSI C78.81 or IESNA-63, set the measurement angles on the goniophotometer so that they match those angles. The goniophotometer's photometer o